Patient Data Module
2030047-001A
Appendix B– Electromagnetic Compatibility
Electromagnetic compatibility (EMC)
Changes or modifications to this system not expressly approved by GE can cause
EMC issues with this or other equipment. This system is designed and tested to
comply with applicable regulation regarding EMC and must be installed and put into
service according to the EMC information stated in this appendix.
WARNING
Use of portable phones or other radio frequency (RF) emitting
equipment near the system may cause unexpected or adverse
operation.
WARNING
The equipment or system should not be used adjacent to, or stacked
with, other equipment. If adjacent or stacked use is necessary, the
equipment or system should be tested to verify normal operation in
the configuration in which it is being used.
Guidance and manufacturer’s declaration –
electromagnetic emissions
The Patient Data Module is intended for use in the electromagnetic environment
specified below. It is the responsibility of the customer or user to assure that the
Patient Data Module is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment – Guidance
RF Emissions
EN 55011
Group 1
The equipment uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
RF Emissions
EN 55011
Class A
Harmonic Emissions
EN 61000-3-2
Class A
The equipment is suitable for use in all establishments other than domestic and
those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Voltage Fluctuations/
Flicker Emissions
EN 61000-3-3
Complies
Summary of Contents for PDM
Page 7: ...2030047 001A Patient Data Module 1 1 1 Introduction ...
Page 13: ...2030047 001A Patient Data Module 2 1 2 Equipment Overview ...
Page 21: ...2030047 001A Patient Data Module 3 1 3 Installation ...
Page 26: ...3 6 Patient Data Module 2030047 001A Installation ...
Page 27: ...2030047 001A Patient Data Module 4 1 4 Service Tool and Configuration ...
Page 38: ...4 12 Patient Data Module 2030047 001A Service Tool and Configuration ...
Page 39: ...2030047 001A Patient Data Module 5 1 5 Maintenance ...
Page 48: ...5 10 Patient Data Module 2030047 001A Maintenance ...
Page 49: ...2030047 001A Patient Data Module 6 1 6 Troubleshooting ...
Page 54: ...6 6 Patient Data Module 2030047 001A Troubleshooting ...
Page 55: ...2030047 001A Patient Data Module 7 1 7 Field Replaceable Units ...
Page 69: ...2030047 001A Patient Data Module A 1 A Appendix A Technical Specifications ...
Page 80: ...A 12 Patient Data Module 2030047 001A Appendix A Technical Specifications ...
Page 81: ...2030047 001A Patient Data Module B 1 B Appendix B Electromagnetic Compatibility ...
Page 90: ...B 10 Patient Data Module 2030047 001A Appendix B Electromagnetic Compatibility ...
Page 91: ......