Field Replaceable Units
2030047-001A
Patient Data Module
7-13
NBP assembly with manifold hose
1. Remove the top housing, main board, NBP hose, coupling and elbow per above
steps.
2. Remove 4 screws from the battery cover.
467A
3. Grasp and slide the battery cover out to the rear.
4. Remove 1 screw under the pump tube connection.
466A
5. Lift off the manifold with pump.
6. Remove the interface gasket.
Front panel board connector
Main board flex pump connector
Battery cover
Summary of Contents for PDM
Page 7: ...2030047 001A Patient Data Module 1 1 1 Introduction ...
Page 13: ...2030047 001A Patient Data Module 2 1 2 Equipment Overview ...
Page 21: ...2030047 001A Patient Data Module 3 1 3 Installation ...
Page 26: ...3 6 Patient Data Module 2030047 001A Installation ...
Page 27: ...2030047 001A Patient Data Module 4 1 4 Service Tool and Configuration ...
Page 38: ...4 12 Patient Data Module 2030047 001A Service Tool and Configuration ...
Page 39: ...2030047 001A Patient Data Module 5 1 5 Maintenance ...
Page 48: ...5 10 Patient Data Module 2030047 001A Maintenance ...
Page 49: ...2030047 001A Patient Data Module 6 1 6 Troubleshooting ...
Page 54: ...6 6 Patient Data Module 2030047 001A Troubleshooting ...
Page 55: ...2030047 001A Patient Data Module 7 1 7 Field Replaceable Units ...
Page 69: ...2030047 001A Patient Data Module A 1 A Appendix A Technical Specifications ...
Page 80: ...A 12 Patient Data Module 2030047 001A Appendix A Technical Specifications ...
Page 81: ...2030047 001A Patient Data Module B 1 B Appendix B Electromagnetic Compatibility ...
Page 90: ...B 10 Patient Data Module 2030047 001A Appendix B Electromagnetic Compatibility ...
Page 91: ......