4-2
Patient Data Module
2030047-001A
Service Tool and Configuration
Overview
The Patient Data Module service tool provides the following:
Device information
Configuration including asset setting, ECG filter configuration, licensing and
software transfer
Diagnostics
Calibration
All service functions for the Patient Data Module are performed through a service
laptop.
Required equipment
The following is a list of equipment to use with the GEHC MS service tool.
If servicing the Patient Data Module using a Solar 8000M/I patient monitor:
PC with Microsoft
®
Windows
®
XP or 2000
™
operating system, a CD-ROM
drive, and a network card
Category 5 crossover cable (2028822-001)
ePort to host interface cable (2017098-002)
Solar 8000M/i patient monitor with PDM service adapter (2028845-002)
Patient Data Module software CD with GEHC MS Service Tool
If servicing the Patient Data Module without a host patient monitor:
PDM Service Tool Kit (2030924-001)
Set up the PC
Connect the PC to the Patient Data Module
There are two ways to connect the PC to the Patient Data Module. One method uses
power from the Solar 8000M/i patient monitor, the other method uses an external
power supply.
Summary of Contents for PDM
Page 7: ...2030047 001A Patient Data Module 1 1 1 Introduction ...
Page 13: ...2030047 001A Patient Data Module 2 1 2 Equipment Overview ...
Page 21: ...2030047 001A Patient Data Module 3 1 3 Installation ...
Page 26: ...3 6 Patient Data Module 2030047 001A Installation ...
Page 27: ...2030047 001A Patient Data Module 4 1 4 Service Tool and Configuration ...
Page 38: ...4 12 Patient Data Module 2030047 001A Service Tool and Configuration ...
Page 39: ...2030047 001A Patient Data Module 5 1 5 Maintenance ...
Page 48: ...5 10 Patient Data Module 2030047 001A Maintenance ...
Page 49: ...2030047 001A Patient Data Module 6 1 6 Troubleshooting ...
Page 54: ...6 6 Patient Data Module 2030047 001A Troubleshooting ...
Page 55: ...2030047 001A Patient Data Module 7 1 7 Field Replaceable Units ...
Page 69: ...2030047 001A Patient Data Module A 1 A Appendix A Technical Specifications ...
Page 80: ...A 12 Patient Data Module 2030047 001A Appendix A Technical Specifications ...
Page 81: ...2030047 001A Patient Data Module B 1 B Appendix B Electromagnetic Compatibility ...
Page 90: ...B 10 Patient Data Module 2030047 001A Appendix B Electromagnetic Compatibility ...
Page 91: ......