Calibration Check (All Ranges)

Note:

The PACE adds the barometric reading to a gauge range to produce an absolute range.

6.1 Procedure

Set the calibrator to measure mode:

1

Connect calibration standard for the pressure range to be checked.

2

Press 

Task

 and select 

Basic

.

3

With the pressure standard connected to the correct pressure port, select 

Measured

 

Pressure

 and press 

Range

 to select the gauge pressure range to be checked.

4

Barometric pressure can be displayed in the status area.

5

Set maximum displayed resolution.

Gauge ranges should be zeroed immediately prior to a calibration check, as follows:

1

Press 

Measured Pressure/Zero

 to zero the selected range.

2

On completion of the zero operation, the display shows Zero completed successfully.

3

Adjust calibration pressure to the first pressure value and wait until this pressure, as 
displayed on PACE, is stable to less than 10ppm (0.001%). Display filtering may be 
required.

4

Compare the pressure value on the calibration standard to the value displayed and 
record the difference.

5

Repeat (3) and (5) for each pressure, as prompted by the PACE.

6

If the recorded difference exceeds the permissible deviation for the selected range, the 
calibrator requires a calibration adjustment for that range. Refer to PACE sales data 
sheets for permissible precision deviation. Precision specification is used if less than 24 
hours since adjustment using the same calibration equipment. If greater than 24 
hours, then the PACE specification is the sum of precision and long term stability.

7

Select the next pressure range for a calibration check.

8

After completing all calibration checks, adjust calibration standard to atmospheric 
pressure.

9

Disconnect calibration standard from the output.

If no further calibration is required, switch off the PACE.

Summary of Contents for PACE5000

Page 1: ...GE Digital Solutions Pressure Automated Calibration Equipment Calibration manual K0450 Revision B PACE5000 PACE6000 PACE1000...

Page 2: ...ons are subject to change without notice GE is a registered trademark of General Electric Company Other company or product names mentioned in this document may be trademarks or registered trademarks o...

Page 3: ...ge Use suitably qualified calibration technicians and good engineering practice for all procedures in this publication Pressure Do not apply pressures greater than the maximum working pressure to the...

Page 4: ...d airspeed i e That is IEC International Electrotechnical Commission IEEE 488 Institute of Electrical and Electronic Engineers standard 488 data in Inch kg Kilogram kts kn knot LCD Liquid crystal disp...

Page 5: ...ets the requirements of all relevant European safety directives The equipment carries the CE mark This symbol on the equipment indicates that the user should read the user manual This symbol on the eq...

Page 6: ...RE APPLYING PRESSURE EXAMINE ALL FITTINGS AND EQUIPMENT FOR DAMAGE REPLACE ALL DAMAGED FITTINGS AND EQUIPMENT DO NOT USE ANY DAMAGED FITTINGS AND EQUIPMENT DO NOT EXCEED THE MAXIMUM WORKING PRESSURE O...

Page 7: ...r the list of selectable pressure units Pressure units Factor hPa Pressure units Factor hPa mbar 1 0 cmH2O 20 C 0 978903642 bar 1000 0 mH2O 20 C 97 8903642 Pa N m2 0 01 kg m2 0 0980665 hPa 1 0 kg cm2...

Page 8: ...es on Calibration 1 5 1 PACE Controller Connection 2 Gauge Reference 2 5 2 PACE Indicator Connection 3 Gauge Reference 3 6 Calibration Check All Ranges 4 6 1 Procedure 4 7 Calibration Adjustment 5 App...

Page 9: ...rement uncertainty of the original calibration standard To preserve uncertainty of the PACE calibration checks and adjustments must be performed using a calibrator uncertainty of less than or equal to...

Page 10: ...open to atmosphere Turn the valve to the closed position Gauge Reference If the pressure standard has a reference connection then connect this to the PACE reference port on the module manifold Otherw...

Page 11: ...essure standard has a reference connection then connect this to the indicator reference port Otherwise the indicator reference port should be open to atmosphere Note Indicator reference port open to a...

Page 12: ...e to the first pressure value and wait until this pressure as displayed on PACE is stable to less than 10ppm 0 001 Display filtering may be required 4 Compare the pressure value on the calibration sta...

Page 13: ...e to be set on the pressure standard and to press OK when the applied pressure is stable Use the numeric keys to enter the precise applied pressure Note The display also shows throughout this procedur...

Page 14: ...K0450 Revision B 6 EN English Appendix A PACE Menus and Screens A 1 PACE Controller Menus Calibration procedure next page...

Page 15: ...PACE Series Calibration Manual EN English 7 K0450 Revision B A 2 PACE Controller Screens...

Page 16: ...K0450 Revision B 8 EN English A 3 PACE Indicator Menus Calibration procedure next page...

Page 17: ...PACE Series Calibration Manual EN English 9 K0450 Revision B A 4 PACE Indicator Screens...

Page 18: ......

Page 19: ......

Page 20: ...www gemeasurement com...

Reviews: