Chapter 13: Advanced Applications
5495975-1EN Rev.9
13-18
© 2013-2017 General Electric Company. All rights reserved.
Move Lateral Positioning Bars
1. Hold the Lateral Positioning Bar and pull the lock release to adjust the vertical placement.
Figure 13-15
Lock release in the unlocked position
2. Move the lateral positioning bar to the desired height and then release, thus the positioning bar will
be locked.
Note:
The maximum weight the lateral positioning bar can support is 55Kg (122 lbs.).
Select Auto Image Paste Protocol
To acquire pasted images, you must define the top and bottom of the coverage area. The system calcu-
lates the number of images required to create the pasted image once the coverage area is defined.
During acquisition, the tube and receptor move into position to expose the top of the coverage area. The
tube and receptor automatically move into position to acquire the exposures. Exposures will always
begin at the top of the coverage area (head) and move downward (towards the feet) in sequence.
Follow this process to conduct an Auto Image Paste exam.
1. For wallstand exams, place the patient positioner in front of the wallstand and secure the floor lock-
Lock Positioner in Place (p. 13-16)
for more information.
2. Select the patient from the Worklist (or add patient if necessary).
3. Select the Auto Image Paste protocol from the Select Protocols screen (
).
a) Select the Image Pasting category.