Chapter 8: System Hardware Overview
5495975-1EN Rev.9
8-9
© 2013-2017 General Electric Company. All rights reserved.
Hand Switch
Exposures are made with the console Hand-switch. The Prep/Expose button on this switch has three
positions: OFF, PREPARE, and EXPOSE (
Figure 8-3
Console hand-switch and holder
Use this procedure to operate the console Hand-switch to prepare and record exposures.
1. Make sure your patient and the console are set up for the procedure.
2. Press the Prep/Expose button to the PREPARE position.
Active components
for this product:
1. PREPARE (target
rotating)
2. EXPOSE
3. COLLIMATOR
LIGHT (Not active
on this product)
Table 8-4
Hand-switch positions
Position
Description
OFF
The OFF position is when no pressure is applied to the Prep/Expose button on top of
the Hand-switch.
PREPARE
PREPARE is the next position on the Hand-switch. When it is partially pressed, it
brings the rotor up to speed and heats the filament. PREPARE also checks the
system interlocks and verifies the system is ready to make an exposure. If you
release the button, it returns to OFF.
EXPOSE
The EXPOSE position is when the button on the Hand-switch is fully depressed. This
produces X-rays that are recorded. Release the Prep/Expose button after the
exposure is completed.
The system beeps as X-rays are produced. The tone ends when the exposure is
terminated or completed. The X-ray exposure indicator on the RCIM also lights up
when X-rays are produced.