Chapter 8: System Hardware Overview
5495975-1EN Rev.9
8-7
© 2013-2017 General Electric Company. All rights reserved.
Figure 8-2
RCIM
Table 8-3
RCIM controls
Control
Description
1. Emergency
Stop button
Immediately powers down the system (including table, OTS,
wallstand, and x-ray tube) and stops image exposure.
To engage: Press the button.
To release: Turn the button clockwise (indicated by the arrows
on the button) until it stops, then release.
Refer to
Chapter 4: General Information-Emergency Stop
for more information.
2. Power On
button
Turns the power ON for the entire system.
Refer to
Chapter 4: General Information-System Start Up and
for more information.
3. Reset button
Shuts down and re-starts the system in the event of software
failure.
Refer to
Chapter 4: General Information-System Start Up and
for more information.
4. Expose Hold
indicator
Lights up when there is some inhibit to taking exposures. Click
the Expose Hold button on the Worklist or Acquisition screens
or the OTS control screen for a list of all current inhibits.
5. Tube
Overheat
indicator
Lights up when the tube is too hot to continue taking
exposures.
Allow the tube to cool until the LED turns off.