Chapter 5 Calibration
Page 113
responsible for providing evidence of PACS monitor calibration.
2.
At the application desktop, press
IMAGE TOOLS.
3.
Click SMPTE to load the SMPTE pattern into the Patient List.
4.
Send the SMPTE pattern to the PACS review workstation
5.
Verify that 5% and 95% squares are equally visible on PACS (
).
Figure 5-2 SMPTE pattern
5.4.2.2 Loading test images
1.
Load the PACS test images onto the acquisition workstation. Use either:
-
A DICOM-formatted CD-ROM with images.
-
Images that are stored on the system in Image Tools.
2.
Verify that the patient name is
GE_DX_Image_Display_Test, GEHC
and the patient ID is
2000002
.
5.4.2.3 Sending no-burn images to the PACS
1.
In the SUIF, press the UTILITIES tab button.
2.
Next to Network Connections, press EDIT.
3.
Select the appropriate network host and press EDIT.
4.
On the Preferences tab, verify that the following parameter is NOT checked:
Apply Burn-on-Send to images when sending to this network host.
5.
In the Images list, highlight
PACS Test Image #1
.