Section 18.0 - Seismic Mounting
18.2.2
Power Distribution Unit
CAUTION
The PDU is very heavy and may present a crush hazard if proper precaution and tools are
not used.
If site specifications require seismic mounting, use 5/8 in. (15mm) bolts, and the seismic brackets
and anchors that shipped with the NGPDU. Refer to
for mounting hole locations, and
mount the PDU so it can be easily removed for service.
Figure 1-79 Seismic PDU Mounting Hole Locations
Seismic Bolt Anchor
Locations, both Sides
SERVICE AREA
INDICATES CENTER OF GRAVITY
13.8
(350)
19.4
(494)
41.8
(1062)
27.6
(700)
35.4
(900)
21.7
(550)
9.3
(235)
SEISMIC FLOOR
MOUNTING HOLES
(not to scale; see note)
Front View
Top View
Side View
Seismic Mounting Holes