Upgrade software
2026419-033E
CIC Pro™
12-9
CAUTION
If a USB memory stick other than the stick containing the reload
image is plugged into the CIC Pro center, you need to remove it
prior to rebooting. Otherwise, the message
Operating system
missing
will be displayed.
Complete the checkout procedures
NOTE
After upgrading the CIC Pro center software from v5.0.x to CIC v5.1.x, the
existing Multimouse configurations are lost. Before proceeding to checkout, you
must re-configure Multimouse on the CIC Pro center and ensure that the
Multimouse functionality regained.
See Perform MultiKM (Multimouse) setup
NOTE
The Log Compression Configuration Utility must be run:
After upgrading to CIC Pro center software v5.1.x
When a CIC Pro center v5.1.x is added to an existing Multimouse group
Initial configuration of any CIC Pro center v5.1 device during new installs
During CIC Pro center v5.1 configuration after software re-image/ reload
process
After activating the installed software and before using the CIC Pro center to monitor
patients, you must verify proper operation of this device in the patient care and
networking environments.
To verify proper operation in the patient care and networking environments, complete
all of the checkout procedures identified in
Summary of Contents for MP100 Series
Page 12: ...x CIC Pro 2026419 033E...
Page 13: ...2026419 033E CIC Pro 1 1 1 Introduction...
Page 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview...
Page 53: ...2026419 033E CIC Pro 3 1 3 Licensing...
Page 60: ...3 8 CIC Pro 2026419 033E Licensing...
Page 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces...
Page 74: ...4 14 CIC Pro 2026419 033E Service interfaces...
Page 75: ...2026419 033E CIC Pro 5 1 5 Installation...
Page 94: ...5 20 CIC Pro 2026419 033E Installation...
Page 95: ...2026419 033E CIC Pro 6 1 6 Configuration...
Page 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures...
Page 208: ...7 22 CIC Pro 2026419 033E Checkout procedures...
Page 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting...
Page 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs...
Page 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive...
Page 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount...
Page 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options...
Page 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance...
Page 279: ...2026419 033E CIC Pro 11 1 11 Reload software...
Page 288: ...11 10 CIC Pro 2026419 033E Reload software...
Page 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software...
Page 298: ...12 10 CIC Pro 2026419 033E Upgrade software...
Page 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility...
Page 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules...
Page 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility...
Page 312: ...C 4 CIC Pro 2026419 033E Device compatibility...
Page 313: ......