9-24
CIC Pro™
2026419-033E
Field replaceable units (FRUs)
8.
Re-assemble the cover on page 9-24
9.
Perform the following checkout procedures:
Perform
Ground continuity test on page 10-6
Perform
Ground (earth) wire leakage current test on page 10-7
.
Perform
Enclosure leakage current test on page 10-8
.
Perform general visual inspection.
See Visual inspection on page 10-3.
Inspect fans and power supply.
Inspect connectors.
Inspect cable insulation.
See External surfaces on page 10-4.
Clean internal components.
See Internal components on page 10-5.
Clean fans.
Clean filters.
Check access to all other care units on page 7-7
.
Verify that the power LED is on after the unit is powered up.
Re-assemble the cover
WARNING
SHOCK HAZARD —Whenever the cover is removed, you can
disrupt internal components and potentially disrupt proper power or
ground connections. The potential for electric shock exists. When
the cover is replaced, perform
Electrical safety tests on page 10-6
FRU checkout procedures on page 7-10
Re-assemble the cover on the desktop
1.
Slide the cover over the chassis.
2.
Replace the eight screws that connect the cover to the sides of the chassis.
3.
If you did replace any parts after opening up the cover, you will have to perform
Electrical safety tests on page 10-6
and
FRU checkout procedures on page 7-10
.
Re-assemble the cover on the rack-mount
1.
Slide the cover into the chassis.
2.
Replace the four screws that connect the cover to the back of the chassis.
3.
If you did replace any parts after opening up the cover, you will have to perform
Electrical safety tests on page 10-6
and
FRU checkout procedures on page 7-10
.
Summary of Contents for MP100 Series
Page 12: ...x CIC Pro 2026419 033E...
Page 13: ...2026419 033E CIC Pro 1 1 1 Introduction...
Page 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview...
Page 53: ...2026419 033E CIC Pro 3 1 3 Licensing...
Page 60: ...3 8 CIC Pro 2026419 033E Licensing...
Page 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces...
Page 74: ...4 14 CIC Pro 2026419 033E Service interfaces...
Page 75: ...2026419 033E CIC Pro 5 1 5 Installation...
Page 94: ...5 20 CIC Pro 2026419 033E Installation...
Page 95: ...2026419 033E CIC Pro 6 1 6 Configuration...
Page 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures...
Page 208: ...7 22 CIC Pro 2026419 033E Checkout procedures...
Page 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting...
Page 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs...
Page 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive...
Page 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount...
Page 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options...
Page 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance...
Page 279: ...2026419 033E CIC Pro 11 1 11 Reload software...
Page 288: ...11 10 CIC Pro 2026419 033E Reload software...
Page 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software...
Page 298: ...12 10 CIC Pro 2026419 033E Upgrade software...
Page 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility...
Page 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules...
Page 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility...
Page 312: ...C 4 CIC Pro 2026419 033E Device compatibility...
Page 313: ......