Chapter 2. Installation
Moisture Target Series 6 Hygrometer User’s Manual
25
2.6.2 Connecting a VeriDri Probe (cont.)
2.
Using the flying leads at the end of the
VeriDri
cable, connect the probe
to the lower terminal block (pins 10, 11, 12 and 13) on the back of the
MTS6
(see Table 1 and Figure 35 and Figure 36 on page 26).
IMPORTANT:
To maintain good contact at the terminal block and to avoid
damaging the pins on the wiring connector, pull the connector
straight off (not at an angle) the terminal block. Then, make
the cable connections while the connector is off the unit.
Finally, after the wiring is complete, push the connector
straight onto the terminal block (not at an angle).
Figure 34: Bottom Connector Removed
Table 1: VeriDri Wiring Connections
Wire Color
Pin Number
Designation
Brown
10
RTN
Black
11
COM –
White
12
COM +
Blue
13
+15V