Programming with PanaView
4-13
March 2008
Checking the Battery and
Backlight
If your
MMS-3
has the battery option, you can use PanaView to check
on the battery status and backlight settings. To check on the battery
from PanaView:
1.
From the “
New Meter Browser”
option, click on the desired
communications port and meter.
2.
From the network tree, double-click on “
Battery/Backlight
.” The
“
Battery/Backlight Status”
window appears, as shown in
Figure 4-11 below.
3.
The window shows the presence (or absence) of a battery, along
with its power source, voltage, charge time and remaining time to
run. To recharge the battery, click on
[Full Charge]
.
4.
The window also shows if the backlight is on and how long the
backlight timeout is. Use the numeric keys to enter a backlight
interval in minutes from 0 to 960 (16 hours) in the text box. If you
enter 0 (zero) the backlight will turn off.
5.
Click on
[Apply]
to apply the settings immediately, and on
[OK]
to
confirm the settings and close the window. To return the window
to its previous settings, click on the
[Cancel]
button.
Figure 4-11: Battery/Backlight Status