March 2008
3-32
Advanced Programming
Setting Error Processing
The
MMS-3
provides Error Processing that enables you to specify
how the recorders and alarms will respond to low and high range
errors.
Low and high range errors occur when an input signal that is within
the capacity of the analyzer exceeds the calibration range of the
probe. The
MMS-3
displays an
UnderRng
message for low range
errors and an
OverRng
message for high range error. Refer to
Range
Error Description
in Chapter 2 of the
Service Manual
for more details
on range errors.
The
MMS-3
outputs will also respond to signal and calibration errors.
When a signal error occurs, the outputs will respond in the same
manner as the high range error has been programmed. When a
calibration error occurs, the outputs will respond in the same manner
as the low range error has been programmed. For more details on
signal and calibration errors, refer to
Signal Error Description
and
Calibration Error Description
in Chapter 2 of the
Service Manual.
You can program separate actions for low and high range errors for
the alarm relays and recorder outputs. The actions are independently
programmable for each channel.
Press the
[PROG]
key to enter the user program.
Note:
If you have already entered the user program, see Figure 3-2
on page 3-48 to navigate to the System Menu.
Be sure the number displayed in the upper right-hand corner of the
screen is the channel you want to program. If not, press the
[CHAN]
key to select the desired channel.
Enter Passcode: XXXX
Enter the passcode
.
Programming Menu 1
Use the arrow keys to move the
brackets to
SYSTEM
and press
[YES].
[SYSTEM] AUTOCAL
`