4-18
MAC™ 800
2031504-159D
Maintenance
9. Disconnect the AD-DC cable from the bottom side of mainboard.
Reassembling the Mainboard Assembly
1. Reconnect the AC/DC cable to the bottom of the new mainboard.
2. Reconnect the battery cable to the bottom of the new mainboard.
3. Replace the new mainboard assembly on bottom cover assembly as
shown.
4. Replace the 10 screws that were removed in step 6 on page 4-17.
5. Reconnect the printer cable to the new mainboard.
6. Reassemble the LCD assembly.
7. Reassemble the top cover assembly.
8. Reassemble the battery assembly.
9. Connect the power cord to AC power.
10. Restore system setups that were saved to the SD card.
Summary of Contents for MAC 800
Page 7: ...2031504 159D MAC 800 1 1 1 Introduction ...
Page 13: ...2031504 159D MAC 800 2 1 2 Equipment Overview ...
Page 18: ...2 6 MAC 800 2031504 159D Equipment Overview Detailed Description Block Diagram ...
Page 19: ...Equipment Overview 2031504 159D MAC 800 2 7 System Architecture Overview ...
Page 22: ...2 10 MAC 800 2031504 159D Equipment Overview ECG Data Flow With Sampling Rates ...
Page 23: ...2031504 159D MAC 800 3 1 3 Troubleshooting ...
Page 50: ...3 28 MAC 800 2031504 159D Troubleshooting ...
Page 51: ...2031504 159D MAC 800 4 1 4 Maintenance ...
Page 82: ...4 32 MAC 800 2031504 159D Maintenance ...
Page 83: ...2031504 159D MAC 800 5 1 5 Parts Lists ...
Page 86: ...5 4 MAC 800 2031504 159D Parts Lists ...
Page 87: ...Parts Lists 2031504 159D MAC 800 5 5 ...
Page 88: ...5 6 MAC 800 2031504 159D Parts Lists ...
Page 89: ...Parts Lists 2031504 159D MAC 800 5 7 ...
Page 90: ...5 8 MAC 800 2031504 159D Parts Lists ...
Page 103: ...2031504 159D MAC 800 A 1 A Technical Specifications ...
Page 108: ...A 6 MAC 800 2031504 159D Technical Specifications ...
Page 109: ...2031504 159D MAC 800 B 1 B Electromagnetic Compatibility ...
Page 115: ...2031504 159D MAC 800 C 1 C Supplies Accessories ...
Page 121: ......