3-26
MAC™ 800
2031504-159D
Troubleshooting
Arrhythmia Mode Power Up Mode
These steps describe how to navigate to the
Main Menu
after powering
on the MAC 800 system when
Arrhythmia
is selected for
Power Up
mode
in
Basic Setup
.
NOTE
To perform system setup functions, log in as a user who is assigned
setup editing privileges.
1. If the
High Security Mode
is enabled, proceed with step a through
step d when prompted for a
User ID
and
Password
; if the password
prompt does not appear, go to step 2.
a. Type your user ID in the
User ID
field.
b. Press the
Enter
key or press the down arrow key on the
trimpad
to move the focus to the
Password
field.
c. Type your password in the
Password
field.
d. Press the
F5
key to select
Login
.
If the barcode reader option is enabled, a window opens
prompting you to scan the patient barcode.
NOTE
If the barcode prompt does not appear, go to step 3.
2. Press
F6
to select
Cancel
.
3. Press
F5
to select
Cancel
.
4. Press
F6
to select
More
.
5. Press
F5
to select
Main Menu
.
Main Screen Power Up Mode
These steps describe how to navigate to the
Main Menu
after powering
on the MAC 800 system when
Main Screen
is selected for Power up mode
in
Basic Setup
.
NOTE
To perform system setup functions, log in as a user who is assigned
setup editing privileges.
1. If the
High Security Mode
is enabled, proceed with step a through
step d when prompted for a
User ID
and
Password
; if the password
prompt does not appear, go to step 2.
a. Type your user ID in the
User ID
field.
b. Press the
Enter
key or press the down arrow key on the
trimpad
to move the focus to the
Password
field.
c. Type your password in the
Password
field.
d. Press the
F5
key to select
Login
.
Summary of Contents for MAC 800
Page 7: ...2031504 159D MAC 800 1 1 1 Introduction ...
Page 13: ...2031504 159D MAC 800 2 1 2 Equipment Overview ...
Page 18: ...2 6 MAC 800 2031504 159D Equipment Overview Detailed Description Block Diagram ...
Page 19: ...Equipment Overview 2031504 159D MAC 800 2 7 System Architecture Overview ...
Page 22: ...2 10 MAC 800 2031504 159D Equipment Overview ECG Data Flow With Sampling Rates ...
Page 23: ...2031504 159D MAC 800 3 1 3 Troubleshooting ...
Page 50: ...3 28 MAC 800 2031504 159D Troubleshooting ...
Page 51: ...2031504 159D MAC 800 4 1 4 Maintenance ...
Page 82: ...4 32 MAC 800 2031504 159D Maintenance ...
Page 83: ...2031504 159D MAC 800 5 1 5 Parts Lists ...
Page 86: ...5 4 MAC 800 2031504 159D Parts Lists ...
Page 87: ...Parts Lists 2031504 159D MAC 800 5 5 ...
Page 88: ...5 6 MAC 800 2031504 159D Parts Lists ...
Page 89: ...Parts Lists 2031504 159D MAC 800 5 7 ...
Page 90: ...5 8 MAC 800 2031504 159D Parts Lists ...
Page 103: ...2031504 159D MAC 800 A 1 A Technical Specifications ...
Page 108: ...A 6 MAC 800 2031504 159D Technical Specifications ...
Page 109: ...2031504 159D MAC 800 B 1 B Electromagnetic Compatibility ...
Page 115: ...2031504 159D MAC 800 C 1 C Supplies Accessories ...
Page 121: ......