2. Disconnect the patient cable from the MAC 800 side panel connector
as shown in the following illustration.
3. Connect the new patient cable to the side panel connector.
4. Perform the applicable checkout procedures.
Replacing Barcode Reader
1. Power off the system and disconnect from AC power.
2. Disconnect the barcode reader from the USB connector on the
MAC 800 rear panel as shown in the following illustration.
3. If only the cable is to be replaced, disconnect the cable from the
barcode reader using the following instructions.
a. Insert an Allen wrench (or straightened paper clip) in the
small hole in the base of the barcode reader.
Summary of Contents for MAC 800
Page 7: ...2031504 159D MAC 800 1 1 1 Introduction ...
Page 13: ...2031504 159D MAC 800 2 1 2 Equipment Overview ...
Page 18: ...2 6 MAC 800 2031504 159D Equipment Overview Detailed Description Block Diagram ...
Page 19: ...Equipment Overview 2031504 159D MAC 800 2 7 System Architecture Overview ...
Page 22: ...2 10 MAC 800 2031504 159D Equipment Overview ECG Data Flow With Sampling Rates ...
Page 23: ...2031504 159D MAC 800 3 1 3 Troubleshooting ...
Page 50: ...3 28 MAC 800 2031504 159D Troubleshooting ...
Page 51: ...2031504 159D MAC 800 4 1 4 Maintenance ...
Page 82: ...4 32 MAC 800 2031504 159D Maintenance ...
Page 83: ...2031504 159D MAC 800 5 1 5 Parts Lists ...
Page 86: ...5 4 MAC 800 2031504 159D Parts Lists ...
Page 87: ...Parts Lists 2031504 159D MAC 800 5 5 ...
Page 88: ...5 6 MAC 800 2031504 159D Parts Lists ...
Page 89: ...Parts Lists 2031504 159D MAC 800 5 7 ...
Page 90: ...5 8 MAC 800 2031504 159D Parts Lists ...
Page 103: ...2031504 159D MAC 800 A 1 A Technical Specifications ...
Page 108: ...A 6 MAC 800 2031504 159D Technical Specifications ...
Page 109: ...2031504 159D MAC 800 B 1 B Electromagnetic Compatibility ...
Page 115: ...2031504 159D MAC 800 C 1 C Supplies Accessories ...
Page 121: ......