4-14
MAC
5500 resting ECG analysis system
Revision E
2020299-020
Troubleshooting: Diagnostic Tests
Battery temperature
Battery charge status
Percent of charge remaining
Battery Charge Test
This test completely discharges the battery, if necessary, then monitors a charge
cycle.
NOTE
This test can take up to 6 hours to run. The “Battery Discharge Test” is a
better indicator of the condition of the battery.
1.
To perform the Battery Charge test, unplug the unit from AC (mains) power.
2.
Select
Battery Charge Test
. The Battery charge test window will appear and the
unit will begin to discharge the battery.
3.
Once the unit has fully discharged the battery, plug the unit back into AC
(mains) power. The battery will begin to charge.
4.
When the battery charge test is complete, go to
Print the Charge Test Results
.
Monitored information, written to internal memory, includes:
Charge rate (in mAH)
Battery temperature
Battery charge status
Percent of charge
Print Charge/Discharge Test Results
Prints the results of the last discharge or charge test to the writer.
Communication Tests
COM Port Loopback Test
The
Communications Port Loopback Test
sends various ASCII characters out the
COM port’s transmit lines and expects the same character to return in it’s receive
lines. Upon completion of the test, the word
Passed
or
Failed
appears, depending on
the results.
For each of the options listed (COM1, COM2, COM3, and COM4) perform the
following steps,
1.
Select an option and press the
Enter
key.
2.
Follow the instructions on screen and install loopback jumpers in the selected
serial port.
3.
Remove the loopback jumpers when the test is complete.
Summary of Contents for MAC 5500
Page 1: ...g MAC 5500 Resting ECG Analysis System Service Manual 2020299 020 Revision E GE Healthcare ...
Page 9: ...Revision E MAC 5500 resting ECG analysis system 1 1 2020299 020 Introduction 1 Introduction ...
Page 10: ...1 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Introduction For your notes ...
Page 55: ...Revision E MAC 5500 resting ECG analysis system 3 1 2020299 020 Installation 3 Installation ...
Page 56: ...3 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Installation For your notes ...
Page 105: ...Revision E MAC 5500 resting ECG analysis system 5 1 2020299 020 Maintenance 5 Maintenance ...
Page 106: ...5 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Maintenance For your notes ...
Page 137: ...Revision E MAC 5500 resting ECG analysis system 6 1 2020299 020 Parts List 6 Parts List ...
Page 138: ...6 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Parts List For your notes ...
Page 175: ...Revision E MAC 5500 resting ECG analysis system Index 1 2020299 020 Index ...
Page 176: ...Index 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 178: ...Index Index 4 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 179: ......