Revision E
MAC
5500 resting ECG analysis system
B-5
2020299-020
Appendix B – Technical Specifications: Technical Specifications
Hi-Res and PHi-Res Signal-Averaged Electrocardiography
Item
Description
Frequency Response/Input
-3 dB @ 0.01 and 250 Hz
Frequency Response/Output
Upper Limit
Lower Limit
250 Hz
0.01, 25, 40 or 80 Hz
Sensitivities
Raw Data and Template
Average Beat
Filtered Signals and Vector
Magnitude
20 mm/mV
20 mm/mV and 50 mm/mV
1 mm/µV
Analysis Sampling Rate
1000 samples per second per channel
Digital Sampling Rate
4000 samples per second per channel
High/Low Pass Filters
Spectral filter using Fast Fourier Transform (FFT)
ADC Resolution
1.22 µV/LSB
Analysis Resolution
0.1525 µV/LSB
Physical
1
Item
Description
Height
9.4 cm (3.7 in) with display closed
Width
38.1 cm (15.0 in)
Depth
35.1 cm (13.8 in)
Weight
6.8 kg (15 lb) without paper
1. without trolley
Environmental
Item
Description
Operating Conditions
Temperature
Relative Humidity
Atmosphere Pressure
10° C to 40° C (50° F to 104° F)
1
20% to 95% RH noncondensing
700 to 1060 hPa
Summary of Contents for MAC 5500
Page 1: ...g MAC 5500 Resting ECG Analysis System Service Manual 2020299 020 Revision E GE Healthcare ...
Page 9: ...Revision E MAC 5500 resting ECG analysis system 1 1 2020299 020 Introduction 1 Introduction ...
Page 10: ...1 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Introduction For your notes ...
Page 55: ...Revision E MAC 5500 resting ECG analysis system 3 1 2020299 020 Installation 3 Installation ...
Page 56: ...3 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Installation For your notes ...
Page 105: ...Revision E MAC 5500 resting ECG analysis system 5 1 2020299 020 Maintenance 5 Maintenance ...
Page 106: ...5 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Maintenance For your notes ...
Page 137: ...Revision E MAC 5500 resting ECG analysis system 6 1 2020299 020 Parts List 6 Parts List ...
Page 138: ...6 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Parts List For your notes ...
Page 175: ...Revision E MAC 5500 resting ECG analysis system Index 1 2020299 020 Index ...
Page 176: ...Index 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 178: ...Index Index 4 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 179: ......