5-6
MAC™ 3500 Resting ECG Analysis System
Revision L
2021337-036
Maintenance:
FRU Replacement Procedures
FRU Replacement Procedures
Disassembly Guidelines
Prior to performing any disassembly procedures, perform the following:
If possible, process any ECGs remaining in storage.
If possible, print out set-up for future reference.
Disconnect the unit from the AC wall outlet and remove the power cord from
the unit.
Remove the battery as described in
If the MAC 3500 system is mounted to the trolley, remove it from the trolley as
described in
“Remove MAC 3500 System From Trolley”
Remove the writer paper.
Take strict precautions against electrostatic discharge damage.
Battery Replacement
1. Press the writer button to open the unit.
2. Slide the battery release button in the direction of the arrow and lift the battery
out.
writer button
12A
Summary of Contents for MAC 3500
Page 9: ...Revision L MAC 3500 Resting ECG Analysis System 1 1 2021337 036 1 Introduction ...
Page 10: ...1 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 19: ...Revision L MAC 3500 Resting ECG Analysis System 2 1 2021337 036 2 Equipment Overview ...
Page 20: ...2 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 25: ...Revision L MAC 3500 Resting ECG Analysis System 3 1 2021337 036 3 Installation ...
Page 26: ...3 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 41: ...Revision L MAC 3500 Resting ECG Analysis System 4 1 2021337 036 4 Troubleshooting ...
Page 42: ...4 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 83: ...Revision L MAC 3500 Resting ECG Analysis System 5 1 2021337 036 5 Maintenance ...
Page 84: ...5 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 123: ...Revision L MAC 3500 Resting ECG Analysis System 6 1 2021337 036 6 Parts Lists ...
Page 124: ...6 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 153: ...Revision L MAC 3500 Resting ECG Analysis System A 1 2021337 036 A Technical Specifications ...
Page 154: ...A 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 162: ...B 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 168: ...C 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 175: ...Revision L MAC 3500 Resting ECG Analysis System Index 1 2021337 036 Index ...
Page 176: ...Index 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 178: ...Index Index 4 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 ...
Page 179: ......