Service Setup
Testing the Acquisition Module
Use the
Acquisition Module Test
to determine if the acquisition board is working
properly.
1.
Open the
Diagnostic Tests
window as described in
page 46
2.
Select
Acquisition Module Test
.
A window similar to the one shown in the following screen opens.
NOTE:
The version displayed depends on the software installed.
3.
Note the test result and press
Cancel
.
4.
If the result of the
Acquisition Module Test
was
Failed
, replace the main PWA
as described in
.
Checking the Battery Status
Use the
Battery Test
to determine the status of the Lithium-Ion battery. You must
perform this test while running on battery power.
1.
Disconnect the device from AC power and wait at least one minute.
2.
Open the
Diagnostic Tests
window as described in
page 46
3.
Select
Battery Test
.
A window similar to the one shown in the following screen opens.
2053535-003C
MAC™ 2000 ECG Analysis System
51