Maintenance
Replacing the Internal Modem
Use the following procedure to replace the internal modem on the PWA.
1.
Locate the internal modem connectors on the PWA, as identified in the following
photograph.
2.
Grasping the edges of the internal modem between your thumb and fingers,
gently seat the internal modem into the connectors on the PWA, as shown
in the following photograph.
3.
Reattach the top cover assembly.
See
“Reattaching the Top Cover Assembly to the Device” on page 81
for
instructions.
4.
Reinsert the battery.
See step
through step
in
“Replacing the Battery” on page 74
for instructions.
5.
Continue with the appropriate functional checkout procedure for this FRU.
See
“Functional Checkout” on page 129
for more information.
PWA
The process to replace the PWA consists of the following procedures:
1.
Processing ECGs in internal storage
2.
Saving system configuration settings
3.
Printing the system setup report
120
MAC™ 2000 ECG Analysis System
2053535-003C