Revision D
MAC
1200 resting ECG analysis system
3-23
2012250-095
System Test Function: Pharma
Pharma
The MAC 1200 device variant is configured in the service mode and any MAC 1200
can be configured as a PHARMA device.The Pharma device variant is designed for
use by Clinical Research Organizations (CROs) in pharmaceutical trials. MAC 1200
units with the Pharma option are configured for specific pharmaceutical trials by
CROs and then sent out to trial sites to collect data.
The Pharma option of the MAC 1200 can be password-protected (HIGH
SECURITY). Password protection is enabled in the system setup.
High Security
NOTE
The
Pharma
and
HIGH SECURITY
options appear only if the device has been
configured as a Pharma device.
High Security provides the option of password-protecting use of the MAC 1200
unit. Any device configured with the Pharma option may enable the High Security
option.
When the High Security option is enabled, users must provide information for
the Technician Name field.
To bypass the logon password, press
+
+
.
Summary of Contents for MAC 1200
Page 10: ...vi MAC 1200 resting ECG analysis system Revision D 2012250 095 ...
Page 11: ...Revision D MAC 1200 resting ECG analysis system 1 1 2012250 095 1 The Basics ...
Page 12: ...1 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 19: ...Revision D MAC 1200 resting ECG analysis system 2 1 2012250 095 2 Functional Description ...
Page 20: ...2 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 45: ...Revision D MAC 1200 resting ECG analysis system 3 1 2012250 095 3 System Test Function ...
Page 46: ...3 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 69: ...Revision D MAC 1200 resting ECG analysis system 4 1 2012250 095 4 Repair Instructions ...
Page 70: ...4 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 79: ...Revision D MAC 1200 resting ECG analysis system 5 1 2012250 095 5 Troubleshooting ...
Page 80: ...5 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 85: ...Revision D MAC 1200 resting ECG analysis system 6 1 2012250 095 6 Adjustment Instructions ...
Page 86: ...6 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 89: ...Revision D MAC 1200 resting ECG analysis system 7 1 2012250 095 7 Service and Maintenance ...
Page 90: ...7 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 101: ...Revision D MAC 1200 resting ECG analysis system 8 1 2012250 095 8 Parts List ...
Page 102: ...8 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 111: ...Revision D MAC 1200 resting ECG analysis system 9 1 2012250 095 9 Specifications ...
Page 112: ...9 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 123: ...Revision D MAC 1200 resting ECG analysis system 10 1 2012250 095 10 Drawings ...
Page 150: ......