2-24
MAC
1200 resting ECG analysis system
Revision D
2012250-095
Functional Description: Interfaces to Peripherals
Electronic Interfaces
RS-232 Interface
The 9-pin sub-D Connector of the RS-232 interface is implemented on the PCB
Control CS_CI directly.
From Version V1.1, the remote start input pin is available on pin 8.
In MAC1200 US devices the modem supply 8 V is available on pin 6
additionally.
Connector denotation:
RS232
Type:
9-pin sub-D, female, Mark II with 4-40 UNC inserts
The function of the individual pins is given in the table below. The definition as an
input/output is seen with reference to PCB Control CS_CI.
RS232
Table 2. RS232 Pin Position
Pin Number
Signal
Input/Output
Function
Definition
1
NC
2
RXDE
Input
RS232 Data In
3
TXDE
Output
RS232 Data Out
4
NC
5
GND
Signal Ground
6
V_MODEM
Output
Modem 8 V
MAC1200 US only
7
RTSE
Output
Request to Send
8
FERNST
Input
Remote Start Input
0: active
9
NC
Summary of Contents for MAC 1200
Page 10: ...vi MAC 1200 resting ECG analysis system Revision D 2012250 095 ...
Page 11: ...Revision D MAC 1200 resting ECG analysis system 1 1 2012250 095 1 The Basics ...
Page 12: ...1 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 19: ...Revision D MAC 1200 resting ECG analysis system 2 1 2012250 095 2 Functional Description ...
Page 20: ...2 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 45: ...Revision D MAC 1200 resting ECG analysis system 3 1 2012250 095 3 System Test Function ...
Page 46: ...3 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 69: ...Revision D MAC 1200 resting ECG analysis system 4 1 2012250 095 4 Repair Instructions ...
Page 70: ...4 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 79: ...Revision D MAC 1200 resting ECG analysis system 5 1 2012250 095 5 Troubleshooting ...
Page 80: ...5 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 85: ...Revision D MAC 1200 resting ECG analysis system 6 1 2012250 095 6 Adjustment Instructions ...
Page 86: ...6 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 89: ...Revision D MAC 1200 resting ECG analysis system 7 1 2012250 095 7 Service and Maintenance ...
Page 90: ...7 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 101: ...Revision D MAC 1200 resting ECG analysis system 8 1 2012250 095 8 Parts List ...
Page 102: ...8 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 111: ...Revision D MAC 1200 resting ECG analysis system 9 1 2012250 095 9 Specifications ...
Page 112: ...9 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 123: ...Revision D MAC 1200 resting ECG analysis system 10 1 2012250 095 10 Drawings ...
Page 150: ......