Perform the system checkout procedure explained
in Chapter 2 before using the Lullaby
TM
Warmer.
3.6
Cleaning Solutions
Use the only cleaning solutions (or chemical equivalents) listed below for cleaning.
Generic Formulation
Maximum Concentration
Sodium Hypochlorite
0.5% Aqueous Solution
Glutaraldehyde
2%
Hydrogen Peroxide
6%
Cavicide®
100% spray
CAUTION:-
Do not use the following cleaners, they will damage the parts being cleaned:
• Isopropyl Alcohol
• Quaternary Ammonium (such as Virex)
• Solvents (such as acetone)
CAUTION:-
Do
not spray cleaning solutions directly on the Lullaby
TM
Warmer.
CAUTION:-
Apply the cleaning solution with a clean sponge or cloth. Do not allow liquids to seep
into the heater housing.
CAUTION:-
Always dry the surfaces with a clean, damp, soft cloth to avoid scratches and remove
any residue of the cleaning solution left after cleaning.
3.7
Cleaning and Disinfecting Individual Components
Patient Probe (Reusable)
CAUTION:-
Avoid placing excessive strain on the probe lead. When cleaning, be careful not to
pull or bend the lead at the probe tip. Always remove the probe from the Lullaby
TM
Warmer by grasping the plug at the control panel. Do not pull on the probe lead.
CAUTION:-
Do not apply cold sterilization or cleaning solutions to the probe connector.
Summary of Contents for Lullaby Warmer
Page 12: ...This page is left blank intentionally ...
Page 139: ...LullabyTM Warmer 2042376 001 Rev G 2012 General Electric Company All rights reserved 113 ...
Page 140: ...114 2042376 001 Rev G 2012 General Electric Company All rights reserved LullabyTM Warmer ...
Page 150: ...124 2042376 001 Rev G 2012 General Electric Company All rights reserved LullabyTM Warmer ...