LOGIQ Totus
Basic Service Manual
10 - 8
Section 10-4 - System Maintenance
Section 10-4
System Maintenance
10-4-1
Preliminary checks
The preliminary checks take about 15 minutes to perform. Refer to the Ultrasound system user
documentation whenever necessary.
10-4-2
Tools, supplies and equipment
lists the suggested tools, supplies and equipment that can be used during the maintenance
tasks.
Table 10-2 System Preliminary Checks
Step
Item
Description
1.
Ask and Listen
Ask the customer if they have any problems or questions about the equipment.
2.
Paperwork
Fill in the top of the EQC inspection form. Record all probes and Ultrasound system options.
3.
Power up
• Turn the Ultrasound system power on and verify that all fans and peripherals turn on.
• Watch the displays during power up to verify that no warning or error messages are displayed.
• Where applicable, confirm that the batteries are charged. If no AC Input present, use the
internal batteries.
4.
Probes
Verify that the Ultrasound system properly recognizes all probes.
5.
Displays
Verify proper display on the monitor and touch panel (where present).
6.
InSite
Where applicable, for Warranty and Contract Customers only:
• Verify that InSite is functioning properly.
• Ensure two-way remote communications.
7.
Review
Error Logs
Where applicable, error logs can be reviewed through system diagnostics.
8.
Diagnostics
Optional.
9.
Presets
Backup all Customer Presets to an appropriate media.
10.
Image Archive
Back up the Image Archive onto appropriate media.
Summary of Contents for LOGIQ Totus
Page 2: ......
Page 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Page 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Page 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Page 643: ......