LOGIQ T
OTUS
Basic Service Manual
8 - 154
Section 8-8 - Replacing Main Console
8-8-24 Replacement of the Unity PWM Fan Assy
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5
8-2-7 "PPE Required During Service" on page 8-6
8-8-24-1
Preparations
1.) Remove the T2 Side Cover R Assy as described in
8-8-8 "Replacement of the T2 Side Cover R
2.) Remove the Rear Cover Mid Assy-T as described in
8-8-10 "Replacement of the Rear Cover Mid
Table 8-86 Manpower, Time, Tools, and PPE
Manpower
Time
Tools
PPE
One person
Approximately 10 minutes
Philps Screwdriver
Cut resistant gloves
Knee Pad