GE HealthCare CONFIDENTIAL
D
IRECTION
5936427
, R
EV
. 1
LOGIQ T
OTUS
Basic Service Manual
Section 8-7 - Replacing OPIO
8 - 101
8-7-11 Replacement of the Trackball assy-U_FLT20UFA-GJ
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5
8-2-7 "PPE Required During Service" on page 8-6
8-7-11-1
Preparations
1.) Remove the OPIO Top Cover Set -T as described in
8-7-2 "Replacement of the OPIO Top Cover
2.) Remove the UOPIO Assy-T as described in
8-7-10 "Replacement of the UOPIO-T Assy" on page
8-7-11-2
Removal Procedure
8-7-11-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in
Table 8-47 Manpower, Time, Tools, and PPE
Manpower
Time
Tools
PPE
One person
Approximately 10 minutes
Philips Screwdriver
Cut resistant gloves
Knee Pad
Step
Corresponding graphic
1.
Disconnect the trackball cable and unscrew 4 screws to remove
the OPIO Laser Color Trackball with Cable from the UOPIO
Assy-T
Summary of Contents for LOGIQ Totus
Page 2: ......
Page 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Page 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Page 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Page 643: ......