Chapter 8 - Replacement Procedures
8-113
8-17-13
Separating DC4D Unit from IO BOX
1) Unscrew 12 screws to remove the IO BOX Cover.
Figure 8-128 Screw points of IO BOX cover
2) Unscrew five screws to remove the DC4D from the MIO Assy. Refer to following figure.
Figure 8-129 Screw points of DC4D Assy
3) Carefully lift up the DC4D unit.