GE H
EALTHCARE
D
IRECTION
5245279, R
EVISION
3
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
8 - 94
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-44
ACPC II Assy
8-2-44-1
Tools
•
Common philips screwdrivers
•
Allen/Unbraco wrench
•
Stubby screwdriver (Flat tip and Cross tip)
8-2-44-2
Preparations
•
Shut down the system and switch off the main breaker.
8-2-44-3
Removal procedure
1.) Remove the AC Input Box Assy. Refer to the
8-2-42 "ARP II Assy 110V/220V" on page 8-90
.
2.) Disconnect 5 connectors. Refer to the Figure 8-139.
3.) Unscrew 5 screws to remove the ACPC Assy. Refer to the Figure 8-140.
Figure 8-139 ACPC in Transbox
Figure 8-140 Screws of ACPC assy