GE H
EALTHCARE
D
IRECTION
5394141, R
EVISION
5
LOGIQ™ P5 S
EVICE
M
ANUAL
8-70
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-27
EMI Cover R
8-2-27-1
Tools
•
Common pilIips screwdrivers
8-2-27-2
Preparations
•
Shut down the system and switch off the main breaker.
8-2-27-3
Removal procedure
1.) Remove the Side Right Cover. Refer to the
8-2-16 "Side Right Cover" on page 8-52
.
2.) Unscrew 9 screws (1-9).
3.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-27-4
Mounting Procedure
Install the new parts in the reverse order of removal.
Figure 8-105 EMI Cover R
Table 8-28 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test
Debrief Script
“Service Manual, Direction
5394141, Rev 1+, Section 8-2-27. Equipment
passed all required tests and is ready for use. “