Care and Maintenance
10-32
LOGIQ e – Basic Service Manual
5461614-100 English Rev. 6
Grounding continuity
Measure the resistance from the third pin of the attachment plug
to the exposed metal parts of the case. The ground wire
resistance should be less than
0.2
ohms. Reference the
procedure in the IEC60601-1.
Figure 10-7. Ground continuity test
DANGER
ELECTRIC SHOCK HAZARD. THE PATIENT MUST NOT BE
CONTACTED TO THE EQUIPMENT DURING THIS TEST.
1. GROUND
PIN
2. OHMMETER
3. LOGIQ
e
4. ACCESSIBLE METAL PART:
• MONITOR HOUSING
• PEAR PANEL CONNECTOR
• ANY CASTER/WHEEL SUPPORT
Summary of Contents for LOGIQ e
Page 5: ...LOGIQ e Basic Service Manual i 3 5461614 100 English Rev 6...
Page 6: ...i 4 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Page 7: ...LOGIQ e Basic Service Manual i 5 5461614 100 English Rev 6...
Page 8: ...i 6 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Page 9: ...LOGIQ e Basic Service Manual i 7 5461614 100 English Rev 6...
Page 10: ...i 8 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Page 11: ...LOGIQ e Basic Service Manual i 9 5461614 100 English Rev 6...
Page 12: ...i 10 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Page 24: ...i 22 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Page 60: ...Introduction 1 36 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Page 80: ...Site Preparations 2 20 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Page 134: ...System Setup 3 54 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Page 208: ...Components and Functions Theory 5 14 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Page 262: ...Replacement Procedures 8 26 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Page 328: ...Care and Maintenance 10 44 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Page 384: ...Docking Cart Setup 11 56 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Page 426: ...Index 4 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Page 427: ...GE...