Preparing the System for Use
LOGIQ
t
α
100 User Manual
2211157–100 Rev 0
19
Circuit Breaker
The Circuit Breaker is located on the rear panel of the
console. If the Circuit Breaker is ON power is provided to
the system, if it is OFF it removes power to the system. It
automatically shuts OFF power to the system if a power
overload occurs.
If a power overload occurs:
1. Switch OFF all peripherals.
2. Switch OFF the main power switch to the console.
3. Reactivate the Circuit Breaker switch.
The Circuit Breaker switch should stay in the ON
position. If the Circuit Breaker does not remain in the ON
position (or trips again):
1. Disconnect the power cord.
2. Call Service immediately.
DO NOT attempt to use the system.
Foot Switch Connection (Optional)
The foot switch, which is the remote
FREEZE device, is
connected to the rear panel of the system. This optional
foot switch may be used in parallel with or as an
alternative to the
FREEZE key to enhance flexibility to
freeze an image. Use only the GE recommended foot
switch.
5
Summary of Contents for LOGIQ A100
Page 18: ...Introduction LOGIQt α100 User Manual 2211157 100 Rev 0 2 This page left blank intentionally ...
Page 54: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 38 This page left blank intentionally ...
Page 74: ...Device Labels 58 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 76: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 60 This page left blank intentionally ...
Page 88: ...Image Display 72 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 146: ...OB LOGIQt α100 User Manual 2211157 100 Rev 0 130 This page left blank intentionally ...
Page 236: ...Cardiology 220 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 248: ...Urology LOGIQt α100 User Manual 2211157 100 Rev 0 232 This page left blank intentionally ...
Page 251: ...LOGIQt α100 User Manual 2211157 100 Rev 0 235 Control Keys Configuration Using Control Keys ...
Page 252: ...Control Keys LOGIQt α100 User Manual 2211157 100 Rev 0 236 This page left blank intentionally ...
Page 298: ...Probes 282 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 344: ...Maintenance 328 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 346: ...OB Tables LOGIQt α100 User Manual 2211157 100 Rev 0 330 This page left blank intentionally ...
Page 400: ...OB Tables 384 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 408: ...Index Index 8 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...