Preparing the System for Use
16
LOGIQ
t
α
100 User Manual
2211157–100 Rev 0
Before the system arrives (cont’d)
S
Take precautions to ensure that the system is
protected from electromagnetic interference.
Precautions include:
S
Operate the system at least 15 feet away from
motors, typewriters, elevators, and other
sources of strong electromagnetic radiation.
S
Operation in an enclosed area (wood, plaster
or concrete walls, floors and ceilings) help
prevent electromagnetic interference.
S
Special shielding may be required if the
console is to be operated in the vicinity of
Radio broadcast equipment.
This medical equipment is approved, in terms of the
prevention of radio wave interference, to be used in
hospitals, clinics and other institutions which are
environmentally qualified. The use of this equipment in
an inappropriate environment may cause electronic
interference. Refer to the LOGIQ
(2139768) and this manual for details. This equipment
can be used in residential areas only under the
supervision of physicians or qualified sonographers.
Environmental Requirements
For proper functioning of the LOGIQ
α
100 system, care
must be taken when it is transported or stored.
Operational
Storage
Transport
Temperature
10
_
to 40
_
C
50
_
to 104
_
F
–10 to 60
_
C
14
_
to 140
_
F
–40
_
to 60
_
C
–40
_
to 140
_
F
Humidity
30 to 75%
non–condensing
30 to 90%
non–condensing
30 to 90%
non–condensing
Pressure
700 to 1060hPa
700 to 1060hPa
700 to 1060hPa
Table 1. Environmental Requirements
.
Summary of Contents for LOGIQ A100
Page 18: ...Introduction LOGIQt α100 User Manual 2211157 100 Rev 0 2 This page left blank intentionally ...
Page 54: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 38 This page left blank intentionally ...
Page 74: ...Device Labels 58 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 76: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 60 This page left blank intentionally ...
Page 88: ...Image Display 72 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 146: ...OB LOGIQt α100 User Manual 2211157 100 Rev 0 130 This page left blank intentionally ...
Page 236: ...Cardiology 220 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 248: ...Urology LOGIQt α100 User Manual 2211157 100 Rev 0 232 This page left blank intentionally ...
Page 251: ...LOGIQt α100 User Manual 2211157 100 Rev 0 235 Control Keys Configuration Using Control Keys ...
Page 252: ...Control Keys LOGIQt α100 User Manual 2211157 100 Rev 0 236 This page left blank intentionally ...
Page 298: ...Probes 282 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 344: ...Maintenance 328 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 346: ...OB Tables LOGIQt α100 User Manual 2211157 100 Rev 0 330 This page left blank intentionally ...
Page 400: ...OB Tables 384 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 408: ...Index Index 8 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...