Section 6-5 - Z-Release Cable and Gas Spring Adjustment
6-5-4-2
Cable Adjustment
1.) From underneath the handle use a 10mm open-end (small “ignition” style) wrench to screw
in
the
cable adjustment screw to tighten the Z-release cable.
2.) If you cannot remove enough of the play in the cable using this adjustment, see
6-5-5
Cable Adjustment at Gas Spring
6-5-5-1
Preparations
1.) If not already performed, Power Off / Shutdown the scanner as described in
2.) Remove the right- and left-side covers of the scanner as described in
(with Bumper) Replacement Procedure” on page 8-5
.
3.) Remove the front cover of the scanner as described in
4.) Remove the black-and-white printer.
5.) Remove the VCR.
NOTE:
The height of the printer and VCR make them difficult to remove without first removing the front cover.
Figure 6-5 Z-Release Cable Adjustment Screw
Underside of Z-Release Lever
Z-Release Cable
Cable Adjustment Screw
Underside of Screw Head
Halfway Between Casting
and Un-squeezed Lever