Customizing Your System
16
-
94
LOGIQ 7 Basic User Manual
Direction 2392206-100 Rev. 1
Dataflow
A dataflow is a set of pre-configured services. When you select
a dataflow, the ultrasound system automatically works according
to the services associated with the dataflow. The Dataflow tab
allows you to select and review information about dataflows. You
can also create, change, and remove dataflows.
Set up dataflows for the services. Workflow MUST be PRIMARY
in a dataflow. Also, Query/Retrieve MUST be the only service in
a dataflow -- also, set it to Hidden so that it cannot be selected
from the Patient menu.
NOTE:
You must be logged on as Administrator to use the Dataflow tab.
Figure 16-66. Dataflow Preset Menu
Table 16-56: Dataflow
Preset Parameter
Description
Name
Select the dataflow from the list.
Direct Store
Select to store data directly to archive (no buffer
storage).
Hidden
Select so that this dataflow does not appear as a
Dataflow on the Patient menu.
Default Dataflow
Select to use this dataflow as the default dataflow
when you start the system.
Role
Select the priority of the service:Primary - first
priority (can be assigned to input, output, or I+O
services). Secondary - second priority (can be
assigned to input, output, or I+O services). Primary
output - allows primary assignment to an output
service even if an I+O service is already defined as
primary. You can have a primary out with an I/O and
a Primary with Input Service only. You cannot have
a primary In and Primary. If you assign a primary
role to two services with the same direction, the
system displays a warning. It prompts you to re-
define one of the services as secondary.
Summary of Contents for LOGIQ 7
Page 4: ...i 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 This page intentionally left blank ...
Page 16: ...i 14 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 28: ...Introduction 1 12 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 140: ...Preparing for an Exam 4 28 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 306: ...Scanning Display Functions 6 42 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 418: ...Abdomen and Small Parts 8 18 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 644: ...Cardiology 10 128 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 688: ...Urology 12 8 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 698: ...Pediatrics 13 10 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 882: ...Customizing Your System 16 106 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 928: ...Probes and Biopsy 17 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 974: ...User Maintenance 18 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 983: ...GE Medical Systems ...
Page 984: ...1 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...