Optimizing the Image
5
-
20
LOGIQ 7 Online Help
Direction 2392536-100 Rev. 1
Map
Description
The system supplies B, M, and Doppler Mode system maps.
Adjusting
To select a map, press
Gray Map
. A map window displays. The
image reflects the map as you go through the selections.
Values
Gray maps gradually change from least contrasty or softest to
most contrasty.
Map values vary by probe, application, and multi frequency
setting. Map values are returned to the factory or user preset
value when you change the following: Probe, Exam Category,
Exam Calcs, New Patient, or Frequency.
Map
Description
A
Assigns a greater amount of gray scale pixel values to bright reflectors in the image. Useful
when imaging abdomen, liver, kidney, OB, pelvic, etc.
B
Assigns a greater amount of gray scale pixel values to bright reflectors in the image. Useful
when the abdomen, liver, kidney, OB, pelvic, etc.
C
Assigns a greater amount of gray scale pixel values to bright reflectors in the image. Useful
when imaging the abdomen, liver, kidney, OB, pelvic, etc.
D
Assigns a lesser amount of gray scale pixel values to bright reflectors in the image, compared
to Maps B and C. Useful when imaging bright carotid plaque reflectors.
E
Assigns a equal amounts of gray scale pixel values to all reflectors in the image. Useful when
imaging tissue with bright reflectors, e.g., cyst with septations or calcifications.
F
Assigns a lesser amount of gray scale pixel values to bright reflectors in the image. Useful
when imaging arteries and grafts.
G
Assigns a lesser amount of gray scale pixel values to weak reflectors in the image than Map F.
Useful when imaging small parts.
H
Assigns an s-shape to gray scale pixel values. Useful when imaging anatomical signals with
less tissue differentiation, e.g., tendon, vein, carotid, thyroid, breast, etc.
I
S-shaped map. This map highlights tissue differentiation for a certain band of signals. Useful
when imaging structures where you want more contrast, e.g., renal, tendon, vein, carotid,
thyroid, breast, etc.
J
S-shaped map. This map highlights tissue differentiation for a certain band of signals. Useful
when imaging structures where you want more contrast, e.g., renal, tendon, vein, carotid,
thyroid, breast, etc.
K
S-shaped map. This map highlights tissue differentiation for a certain band of signals. Useful
when imaging structures where you want more contrast, e.g., renal, tendon, vein, carotid,
thyroid, breast, etc.
L
S-shaped map. This map highlights tissue differentiation for a certain band of signals. Useful
when imaging structures where you want more contrast, e.g., cardiology.
M
Assigns a lesser amount of gray scale pixel values to weak reflections in the image than Map
C. Useful when imaging abdomen and liver.
Summary of Contents for LOGIQ 7
Page 4: ...i 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 This page intentionally left blank ...
Page 16: ...i 14 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 28: ...Introduction 1 12 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 140: ...Preparing for an Exam 4 28 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 306: ...Scanning Display Functions 6 42 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 418: ...Abdomen and Small Parts 8 18 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 644: ...Cardiology 10 128 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 688: ...Urology 12 8 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 698: ...Pediatrics 13 10 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 882: ...Customizing Your System 16 106 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 928: ...Probes and Biopsy 17 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 974: ...User Maintenance 18 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 983: ...GE Medical Systems ...
Page 984: ...1 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...