MA-003 – MULTITRANS User Guide – Rev 3.0
18-Jan-16
Page 22 of 30
7.2
Sampling Process
Open the door of the product and observe the operating mode on the LCD panel. The
flowchart shown in Figure 7—5 and the accompanying steps summarise the process to
obtain a manual oil sample.
Figure 7—5: Manual oil sampling flowchart
Open sampling
valve and take oil
sample
Close sampling
valve
Remove male fitting
and valve assembly
Replace quick-
connect body
protector
Standby mode
Measurement
mode
Press
Stop
button
Oil Drains
Standby mode
Attach male fitting
and valve assembly
with valve closed
Press
Manual Sample
button x1,2 or 3 to
select oil circuit
Wait at least
several minutes for
oil to purge…
Press
Manual Sample
button
Press
Manual Sample
button
Remove quick-
connect body
protector