MA-025
–
DGA 900 Operator Guide
–
Rev 2.0
12-Apr-19
Page 59 of 92
Table 6-5 lists the Oxygen sensor error codes.
Table 6-5: Oxygen sensor error codes
Code Error
Note
0
Oxygen sensor measurement has finished with no error.
1
System error. Internal state. Initial state of state machine.
2
System error. Internal error. Oxygen sensor thread is busy.
3
Oxygen sensor is disabled.
4
Measurement cancelled by user.
5
System error. Measurement just started.
6
PGA measurement has finished before we achieve state 6. We could not
run Oxygen sensor measurement.
7
In attempt to write -1 to PGA Oxygen Saturation register, writing fails
because the read value is not the same as which has been written.
8
In attempt to clear EE (Electronically Erasable) Store bit in Oxygen
sensor Control register, writing fails due to a communication error.
9
PGA Oil Temperature reading fails due to a communication error.
10
Oxygen sensor Oil temperature writing fails due to a communication
error.
11
Oxygen sensor (Dissolved Oxygen) DO Trigger command writing fails
due to a communication error.
12
Oxygen sensor Status DO_DATA_RDY is not set.
13
Oxygen sensor DO Error Code reading fails due to a communication
error.
14
Oxygen sensor Oxygen reading fails due to a communication error.
15
PGA Oxygen Saturation writing fails due to a communication error.
16
Reading PGA state fails due to a communication error.
17
PGA Transfix state is greater than 34. We lose time to write Oxygen
Saturation and measurement result may have an error.
18
Oxygen sensor DO Error Code register is not null. Oxygen sensor has an
error during measurement.
6.15.2
Other Notifications
If the onboard MicroSD card has exceeded 80% of its storage capacity, the SD card icon
displays. Click the icon to reveal the message as shown in Figure 65. The MicroSD card
should be replaced as soon as possible.
Figure 65: MicroSD card memory usage