Chapter 4. Environmental Requirements
4.1 Relative Humidity and Temperature
NOTICE
Avoid extremes in temperatures
Innova system room climate requirements – relative humidity and temperature. Individual products or
components are classed by their installation area for in use see
Table 4-1 In use - Humidity and
,
Individual products or components are classed separately for Storage, see
Handling Requirements on page 25
.
NOTE
Due to the differential in temperature / humidity between the clinical room and the room
below, the floor creates condensation within the gantry / patient table resulting in part
failure or rusting inside the gantry / patient table.
For systems that are planned to be installed on higher level (second floor or above), the
temperature and humidity of the rooms that are directly below the gantry room should be
maintained same as the exam room requirement. Failure to do so will void the equipment
warranty. Avoid above grade installations where the temperature is high in the area below
the gantry / table cables exit if at all possible.
Table 4-1 In use - Humidity and Temperature
INSTALLA-
TION ROOM
OF PRODUCT
OR COMPO-
NENT
RELATIVE HUMIDITY (NON-
CONDENSING)
TEMPERATURE
IN-USE
IN-USE (See Note (2a))
RECOMMENDED (See Note
(2b))
MIN
MAX
MIN
MAX
MIN
MAX
Examination
room
30%
70%
+15°C +59°F
+32°C +90°F
Design for Patient/ Operator
Comfort
Technical
room (See
Note (1))
30%
75%
+10°C +50°F
+32°C +90°F
+13°C +55°F
+25°C +77°F
Technical
room with
Fluoro UPS
optional
30%
75%
+20°C +68 °F
+25°C +77°F
+20°C +68 °F
+25°C +77°F
Control room
30%
75%
+15°C +59°F
+35°C +95°F
+20°C +68°F
+25°C +77°F
Environmental Requirements
5750182-1EN 3
Innova
TM
IGS 6
129
Summary of Contents for Innova IGS 6
Page 9: ...Page intentionally left blank...
Page 57: ...Figure 2 11 Omega V Table dimensions Equipment Requirements 46 InnovaTM IGS 6 5750182 1EN 3...
Page 61: ...Figure 2 14 Gas box outlets Omega Table Equipment Requirements 50 InnovaTM IGS 6 5750182 1EN 3...
Page 69: ...Figure 2 22 C2 Cabinet dimensions Equipment Requirements 58 InnovaTM IGS 6 5750182 1EN 3...
Page 147: ...Page intentionally left blank Environmental Requirements 136 InnovaTM IGS 6 5750182 1EN 3...
Page 158: ...Figure 5 6 PDB Schematic CE 1 2 Electrical Requirements 5750182 1EN 3 InnovaTM IGS 6 147...
Page 159: ...Figure 5 7 PDB Schematic CE 2 2 Electrical Requirements 148 InnovaTM IGS 6 5750182 1EN 3...
Page 161: ...Figure 5 9 PDB Schematic UL 1 2 Electrical Requirements 150 InnovaTM IGS 6 5750182 1EN 3...
Page 194: ...InnovaTM IGS 6...