2 Mounting Requirements
2.1 Floor Loading and Recommended Mounting Methods
See
. To obtain floor loading and recommended mounting methods for components
not specified in
, refer to the appropriate component Pre-Installation Manual listed
in
Basic Innova LC System Compatibility.
Table 5-2:
DIMENSIONS MM (INCHES)
PRODUCT OR
COMPONENT
NET WEIGHT
KG (LBS)
WIDTH
DEPTH
HEIGHT
LOAD
BEARING
AREA MM
(INCHES)
WEIGHT/
OCCUPIED
AREA
MOUNTING METHOD
Liberty LC
Positioner
750 (1655)
See Illustrations INNOVA LC
POSITIONER DIMENSIONS:
•
Side view
•
Top view
•
Front view
in
Dimension Drawings
Circle
diameter 600
(23.62)
Recommended:
•
Through-Bolts
(12)
Alternates:
•
On Grade 5/8
in, Anchors
(12)
•
Above
Grade 3/4
in. Anchors
(12)
See
,
and
590 (1300) See
note 1
See illustrations in
Dimension
Drawings
•
Omega IV
•
Omega V
571.5x429
(22.5x16.9)
2410 kg/m2
(492.3 lb/ft2)
Table
900 (1983) See
note 1
Tilt
750x600
(30x24)
2000 kg/m2
(409 lb/ft2
Same as Innova
LC Positioner See
IMPORTANT NOTICE
below.
IMPORTANT NOTICE
The Table baseplate or the System baseplate are mandatory to install the table (patient support). The Table must never be installed on grade.
C2 Cabinet
299 (659)
See Illustration C2 Cabinet Dimensions
in
Dimension Drawings
600x900
(23.63x35.44)
478 kg/m2 (98
lb/ft2)
C1 Cabinet
506 (1115.5)
See Illustration C1 Cabinet Dimensions
in
Dimension Drawings
600x900
(23.63x35.44)
838 kg/m2
(172 lb/ft2)
COOLIX 4000
204 (450) See
note 3
556 (21.9)
See note 2/3
830 (32.7)
See note 3
1090 (43)
See note 3
Four casters
Not applicable
Detector Chiller
Thermo-Tek
13.6 (30)
See Illustration Detector Chiller
Thermo–Tek Dimensions in
Dimension Drawings
283x351
(11.14x13.82)
Detector Chiller
Thermo-Con
14.6 ( 32.2)
See Illustration Detector Chiller
Thermo–Con Dimensions in
Dimension Drawings
13.5x11.14
Fluoro UPS UL
(optional)
530 (1169)
690 (27.0)
800 (31.5)
1820 (70.1)
975 kg/m2
(200 lb/ft2)
Ground -mounted
Fluoro UPS CE
(optional)
480 (1059)
680 (26.7)
800 (31.5)
1450 (57.1)
883 kg/m2
(181 lb/ft2)
Ground -mounted