Main and Extended Menus
Part 17997
Rev. 6, March 2011
4-23
Note: In the versions 2.1 and prior of the H201Ti embedded software, in the Call OnAlarm
mode, the telephone number set in the modem’s register 0 is dialed automatically at
midnight. This function does not exist in version 2.2.
4.5.6.5
H201Ti ID
and
PowerStat. ID
The
H201Ti ID
parameter is used to set the Hydran 201T
i
’s identification number. An
exclusive identification number must be assigned to each H201T
i
in a local network.
The
PowerStat. ID
parameter is used to set the power station’s identification number. The
same power station identification number must be assigned to all H201T
i
’s in a local
network, but
each local network must have a different power station identification number
.
The default values of the
H201Ti ID
and
PowerStat. ID
parameters are therefore usually
modified. Furthermore, the H201T
i
and power station’s identification numbers must
correspond to those assigned to the same H201T
i
’s and power station in the Hydran Host
software.
Note: The identification number of each H201Ti must be recorded during the configuration
of the local network.
4.5.7
Relays/Analog Submenu (Extended Menu Only)
The
Relays/Analog
submenu is used to:
• Read the state of the alarm relays:
On
or
Off
.
• Set the operation mode of the three alarm relays.
• Select the triggering conditions of the fail alarm (
FaultTrig.
).
• Read the state of the analog output and TDM signal (in %).
• Set the operation mode of the analog output and TDM signal.
4.5.7.1
Gas H Relay
•
RelayH State
: Displays the status of the High alarm relay. The two possible states are
explained in Section 10.2.2 on page 10-4.
CAUTION
WARNING
The identification number of at least one H201Ti in a local network
must be set between 1 and 5 (inclusively).
Summary of Contents for Hydran 201i
Page 1: ...GE Energy Services Digital Energy Hydran Instruction Manual 201i System ...
Page 6: ...Instruction Manual vi Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 8: ...Instruction Manual viii Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 22: ...Instruction Manual xxii Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 36: ...Chapter 1 Introduction 1 10 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 114: ...Chapter 5 Before Installation 5 16 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 143: ...Installation Part 17997 Rev 6 March 2011 6 29 Figure 6 9 Installing the CPU Module ...
Page 236: ...Chapter 12 Troubleshooting 12 20 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 244: ...Chapter 13 Maintenance 13 8 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 254: ...Appendix B Mechanical Drawings B 2 Rev 6 March 2011 Part 17997 Figure B 3 1 5 In Adaptor ...
Page 255: ...Hydran 201Ti Intelligent Transmitter Part 17997 Rev 6 March 2011 B 3 Figure B 4 1 In Adaptor ...
Page 284: ...Appendix G Unit Conversions G 2 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 298: ...Appendix I Glossary I 12 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 299: ......