10.2
ALARM RELAYS
The Hydran 201T
i
and the Hydran 201C
i
-1 both have three alarm relays, in other words
one relay per alarm level:
• One gas High alarm relay
• One gas High-High alarm relay
• One fail alarm relay, referred to as
System
in the H201T
i
’s menus
Note: The Hydran 201Ci-C does not have alarm relays. The H201Ci-1 alarm relays are
more accessible than the H201Ti’s relays. Generally, the H201Ci-1 alarm relays are the
ones connected to an alarm panel or SCADA system.
10.2.1 Alarm Contacts
Each relay has:
• A normally open (NO) contact
• A normally closed (NC) contact
• A common contact (Type C)
The physical form of these contacts is as follows:
• Hydran 201T
i
: See the Alarm Contacts 9-Terminal Block (item 12 in Figure 2-5 on
• Hydran 201C
i
-1: See the terminals 7 to 15 of the Main Terminal Block (item 6 in
• Hydran 201C
i
-C: Does not have alarm relays and thus no alarm contacts.
10.2.2 Status of Alarm Relays
The gas High and High-High alarm relays can only be in two states:
• On (coil energized, alarm state LED’s turned on)
• Off (coil de-energized, alarm state LED’s off)
Note: The relays of the Hydran 201Ci Controller always adopt the state of the
Hydran 201Ti’s corresponding relay (except if the supervisory link is broken, in which case
Summary of Contents for Hydran 201i
Page 1: ...GE Energy Services Digital Energy Hydran Instruction Manual 201i System ...
Page 6: ...Instruction Manual vi Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 8: ...Instruction Manual viii Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 22: ...Instruction Manual xxii Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 36: ...Chapter 1 Introduction 1 10 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 114: ...Chapter 5 Before Installation 5 16 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 143: ...Installation Part 17997 Rev 6 March 2011 6 29 Figure 6 9 Installing the CPU Module ...
Page 236: ...Chapter 12 Troubleshooting 12 20 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 244: ...Chapter 13 Maintenance 13 8 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 254: ...Appendix B Mechanical Drawings B 2 Rev 6 March 2011 Part 17997 Figure B 3 1 5 In Adaptor ...
Page 255: ...Hydran 201Ti Intelligent Transmitter Part 17997 Rev 6 March 2011 B 3 Figure B 4 1 In Adaptor ...
Page 284: ...Appendix G Unit Conversions G 2 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 298: ...Appendix I Glossary I 12 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Page 299: ......