Page 55
6.1.2
Left Access Panel Removal
The left access panel needs to be removed for many
of the procedures shown in this section. Instructions
follow.
TOOL LIST
Standard toolkit
SUPPLIES
None
SAFETY CONSIDERATIONS
X-ray shielding contains lead and lead compounds,
chemicals known (to the State of California) to cause
cancer and other reproductive harm. Wear gloves
when handling shielding and wash hands after
handling.
PRECONDITIONS
Machine locked out. See 6.1.1 for details.
REMOVE
1.
Using Phillips screwdriver, remove eleven screws
and washers securing left panel to unit.
View from rear