Section 4-5 - Backup and Restore Database, Preset Configurations and Images
4-5-6
Archiving Images
4-5-6-1
Save Image Archive
NOTE:
A backup of the Image Archive always contains the Patient Archive (database containing patient
demographic data and measurements)
+ images
of the selected exams.
1.) Press the
UTILITIES
key on the control panel.
2.) In the “Utilities” menu touch SYSTEM SETUP to invoke the setup desktop on the screen.
3.) On the left side of the screen select BACKUP and then click the
IMAGE ARCHIVE
tab.
4.) Click the SAVE button.
CAUTION
!! CAUTION:
It is recommended to
Save Image Archive
(
) and
Backup the Full
System Configuration
(
)
once a week.
Figure 4-35 System Setup - Backup - IMAGE ARCHIVE page
Figure 4-36 Image Archive Save - preparations
[Save] button
[Load] button
click [Next] button
[Advanced] button
choose archiving
span time
check mark if desired