Chapter 8 - Replacement Procedures
8-157
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
8-7-11
MPB TO T-FEPS - FRU Cable Replacement Procedure
8-7-11-1
Tools
•
Appropriate Phillips screwdriver
•
Flat-head screwdriver
8-7-11-2
Time Required
15 min
8-7-11-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
.
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-7-11-4
MPB TO T-FEPS Cable Removal Procedure
1.) Perform MPB removal procedure.
2.) Disconnect the following cables connected to BE module:
•
MPB to BIB
•
MPB to BEP (white)
•
Cockpit to BEP split cable (green and black)
•
Printer cable (yellow)
•
Figure 8-164 Disconnecting Cables Connected to BE Module