Chapter 8 - Replacement Procedures
8-139
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
8-7-6
ON/OFF Switch Cable Replacement Procedure
8-7-6-1
Tools
Flat screwdriver.
8-7-6-2
Time Required
10 min
8-7-6-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
.
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-7-6-4
ON/OFF Switch Cable Removal Procedure
1.) Remove all accessories.
2.) Remove the following covers: Lower Front eTower, Left side eTower, and Right Side eTower,
3.) Remove the battery packs to gain access to the MPB module.
4.) Unscrew the two mounting screws and remove the MPB module.
FRU Part #
Refer to
.
•
Accessories - Replacement Procedures
•
RS Probe Cover Removal Procedure
•
Lower Front eTower Cover Removal Procedure
•
Left Side eTower Cover Removal Procedure
•
Right Side eTower Cover Removal Procedure
•
Battery Module Removal Procedure
Figure 8-141 Accessing ON/OFF Switch Cable
ON/OFF Switch Cable