Blue
P1
Green P2
Brown P3
6600-1097-603
Connector
Figure 5-38
Connector Drawing
5.5.5.4 Removing the Old Harness from the Cable Carrier Channel
1. Open the first 4 exposed cable guard latches at the cable carrier channel end.
Figure 5-39
Opened Cable Guard Latches
2. Unsnap the links with open covers leaving the end riveted link and 2 additional links on the cable carrier
channel.
3. Remove the 4 cable ties securing the harness to the channel. (Make note of routing and orientation).
4. Remove and discard the harness. You may want to save the cable carrier links to use as spares.
5.5.5.5 Installing the New Harness on the Cable Carrier Channel
1. On the replacement harness open the first 4 cable guard latches at the cable carrier channel end.
2. Remove the end link and the 2 adjoining links and set aside.
© 2001 by Datex-Ohmeda, Inc.. All rights reserved.
6600-0343-000 104
109
Chapter 5: Repair Procedures
Summary of Contents for Giraffe OmniBed
Page 10: ...x 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Table of Contents ...
Page 28: ...xxviii 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved About this Manual ...
Page 48: ...20 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Chapter 2 Service Checkout ...