2-1
Chapter 2: Installation and Service Checkout
2.1 Installation
The OmniBed is shipped with the canopy in the locked down position. Before the canopy can be raised the
rail shipping locks must be released. They are located in both sides of the OmniBed near the outside bottom
of the uprights. An orange tear-away label marks their location. The lock consists of a socket head cap
screw in a slot. Tightening the screw disengages the screw head from the slot and releases the lift rail.
Using the 4 mm hex key provided with the device, turn the screw clockwise about 8 rotations until the screw
securely seats in its hole. Remove the tear away label. Perform all steps in Chapter 2 Installation and Service
Checkout including the pre-use checkout as indicated.
For units shipped with the Servo Oxygen option, the Oxygen Cells are shipped separately and must be
installed at the time of product installation. Be sure to complete the procedures in “Installing Oxygen
Sensors” on page 7-60 before performing the Servo Oxygen portions of the Service Checkout procedures in
this section of this document.
When mounted to the Carestation unit, electrically powered accessories must be plugged in using the IEC
accessory outlets provided on the back of the Carestation unit. During installation, it is necessary to remove
the plug retaining bracket. Be sure to install the plug retaining bracket as shown in “Power Cord Condition”
on page 2-2 of this document.
2.2 Service Checkout
Note:
All steps required to complete the pre-use checkout are included in the Service Checkout
Procedures.
2.2.1 Mechanical Checks
The following procedures verify that the device is operational and ready to be used.
!
Warning:
Do not perform the service checkout procedure while the patient occupies the device.
!
Warning:
Complete the service checkout procedure section of this manual before putting the
device into operation. If the equipment fails any portion of the checkout procedure it must be
removed from use and repaired.
!
Warning:
The use of accessories, replacement parts, or power cords other than those specified
by the manufacturer may affect the performance of the device and could result in damage to
the device or unsafe operating conditions.
!
Warning:
I
nsulation on electrical wiring can deteriorate with age. Check for brittle or
deteriorated insulation on power cord and all other electrical wiring.
Summary of Contents for Giraffe OmniBed Carestation CS1
Page 4: ...RH 2 Service Manual ...
Page 24: ...Service Language Disclaimer viii Service Manual ...
Page 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Page 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Page 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Page 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Page 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Page 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Page 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Page 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Page 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Page 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...