7.18 Servo Controlled Oxygen Service Procedures
7.18.1 Installing Oxygen Sensors
.
1.
Slide the drawer to one side and using a 3 mm hex key, loosen the captive screws in the chassis cover
sensor housing door and swing the door down to access the sensor housing.
2.
Using a 3 mm hex key, loosen the two M4 socket head screws that secure the sensor housing cover,
and remove the cover.
3.
Remove the old oxygen sensors. Replace with two new sensors (2084419-001).
Note:
The date on the sensor is a discard date; the sensor should be removed from service on this date. All
sensors are shipped from GEHC at least 12 months prior to this date; sensors should be installed
immediately to ensure maximum operating life.
4.
Reinstall the sensor housing cover and close and secure the door in the chassis cover.
5.
Perform the Pre-use Checkout.
7.18.2 Replacing the Vent Screen
1.
Remove rotating mattress tray, tilt platform and upper pan.
2.
Remove the vent cover by turning it counter clockwise.
3.
Remove and discard the ring shaped vent filter screen from the chassis vent.
4.
Install a new vent screen and reinstall the vent cover.
5.
Perform the Pre-use Checkout.
FIGURE 7-54. Installing Sensors
Oxygen sensor
Sensor housing cover
Sensor housing cover
Captive screw
Chassis cover
Sensor housing door
Summary of Contents for Giraffe OmniBed Carestation CS1
Page 4: ...RH 2 Service Manual ...
Page 24: ...Service Language Disclaimer viii Service Manual ...
Page 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Page 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Page 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Page 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Page 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Page 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Page 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Page 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Page 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Page 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...